Specifications

Specifications

Hardware Specifications

General Specifications

Measurement principle Laser Diffraction Method
Measurement range

 
SALD-MS23: 17nm(0.017μm) to 2500μm
SALD-BC23: 17nm(0.017μm) to 400μm
SALD-HC23: 30nm(0.030μm) to 280μm
SALD-DS5: 300nm(0.3μm) to 2500μm

 Note 1: The measurement range varies according to the shape etc. of the particle.

Measurement Unit: SALD-2300

Light source Red Semiconductor laser (Wavelength 680 nm)
Light detector Detector elements for UV semiconductor laser
Total 84 elements (78 forward, 1 side, 5 back)
System Compliance Class 1 Laser Product, CE compliant
Required power supply 115 or 230 VAC as ordered 100 VA
Dimensions & weight W680mm×D280mm×H430mm, 31kg
Operating Environment Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation)

 Note 2: Reference sample and USB cable (2 m) supplied as standard

Sampler: SALD-MS23

Dispersing bath Capacity: 100~280cm3
Sonicator Frequency about 32 kHz, output about 40 W
Liquid Pump Radial pump, maximum flow rate 2000cm3/min
Liquid Pump Material Stainless (SUS 304, SUS 316), Tetrafluoroethylene (PTFE),
Perfluoroelastmor (FEP) or Kalrez, Thermoflon Pascal (inside)
Liquid Supply Pump Diaphragm pump, maximum flow rate 750cm3/min
Liquid Supply Pump Material Tetrafluoroethylene, polyvinyl dene fluoride
Flow Cell Quartz glass
Required power supply 115 or 230 VAC as ordered, 200 VA
Dimensions & weight W390mm×D520mm×H430mm,18kg
Operating Environment Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation)

 Note 3: USB cable (2 m) supplied as standard

Batch Cell: SALD-BC23

Cell Material Quartz glass
Required Liquid Volume Approx. 12 cm3
Stirrer Mechanism Up-and-down movement of blade
Dimensions & weight W100mm×D120mm×H140mm, 0.8kg
Operating Environment Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation)

High-concentration Sample Measurement System: SALD-HC23

Cell Material Borosilicate glass
Required Liquid Volume Approx. 0.15 cm3
Dimensions & weight W20mm×D100mm×H9mm, 0.2kg
Operating Environment Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation)

Cyclone Injection Type Dry Measurement Unit: SALD-DS5

Sample suction types Cyclone type / One shot type / Hand shot type
Suction nozzle Can be selected from 3 types
Sampling Unit Specifications
System Cyclone type
Communication method USB (PC control)
Required power supply 115/230VAC(±10%), 100VA, 50/60Hz (excluding dust collector and compressor)
Dimensions & weight W 240×D 310×H 210 mm, 10 kg
Operating Environment Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation)
Pressure Regulator Specifications
Primary pressure 0.6 to 0.8 MPa
Secondary pressure 0.05 to 0.5 MPa
Filtration rating Removing particles of 5 μm or larger
Connection to air source Tube of 6 mm outside diameter
Communication method USB (PC control)
Required power supply 115/230VAC(±10%), 100VA, 50/60Hz (excluding dust collector and compressor)
Dimensions & weight W 130×D 223×H 233 mm, 3 kg
Operating Environment Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation)

Requirements for compressor and dust collector

Compressor Output 0.4 kW

 
Minimum pressure Approx. 7 kgf/cm2(approx. 0.69 MPa)
Air discharge 45 L/min
Tank volume Approx. 30 L
Dust collector
 
Type Vacuum cleaner (Paper pack type)
Dust collecting efficiency More than 99% for 0.3 μm particles
Capacity 2.0 m3/min or more
Vacuum Approx. 2000 mm Aq or lower
Suction hose diameter Approx. 32 mm

Software Specification

WingSALD II

Output Conditions
Particle Size (μm) Divisions Fixed 51 or 101 divisions : User settable 51 divisions
Particle Amount (%) Divisions Fixed 51 divisions : User settable 51 divisions
Distribution Basis Count, length, area, or volume
Expression of Cumulative Distribution Oversized or undersized
Expression of Frequency Distribution q, q / Δ×, q / Δlog ×
Smoothing Levels 10 levels
Distribution Function Fitting Rosin-Rammler distribution, logarithmic Gaussian distribution
Data Shifting ±10 levels
Report Function Single data sets (6 templates), overlaid data (5 templates), statistical data, time-series data, or 3D data can be selected and output using batch processing
Data Analysis Functions
Scattering Angle Evaluation Function Evaluates scattering characteristics within micro angle regions for samples such as optical films and sheets.
Data Emulation Functions Emulates measurement results from other instruments and measurement principles, using SALD series measurement results.
Mixture Data Simulation Function Simulates particle size distributions using any mixture ratio of multiple particle size distributions.
Data Connection Function Combines two particle size distributions with different measurement ranges at any particle size point to create a single particle size distribution.
Continuous Measurement Function Continuously measures changes in particle size distributions and particle diameters over time, at intervals as short as one second, and saves the results.
*1: Windows is a registered trademark of Microsoft Corporation in the United States and other countries.
*2: Other company names and product names mentioned in this document are registered trademarks or trademarks of their respective companies.
*3: The TM and ® symbols are omitted in this document.

* Appearance and specifications are subject to change without notice.

This page may contain references to products that are not available in your country.
Please contact us to check the availability of these products in your country.